DPG Phi
Verhandlungen
Verhandlungen
DPG

Erlangen 2001 – wissenschaftliches Programm

Bereiche | Tage | Auswahl | Suche | Downloads | Hilfe

HK: Physik der Hadronen und Kerne

HK 38: Instrumentation und Anwendungen VI

HK 38.4: Vortrag

Mittwoch, 21. März 2001, 17:30–17:45, E

A 90O-Compton polarimeter utilizing silicon drift detectors. — •Jan They, Gerd Buschhorn, Rainer Kotthaus, and Dmitry Pugachev — Max-Planck-Institut f"ur Physik (Werner-Heisenberg-Institut),F"ohringer Ring 6, 80805 M"unchen

A compact Compton polarimeter to be used for energy resolved linear polarization analysis of large angle Parametric X-radiation (PXR) in the energy range below 10 keV has been designed, built and calibrated with monochromatized synchrotron radiation [1]. The instrument employs 4 thermoelectrically cooled silicon drift detectors to determine the linear polarization vector by measuring the azimuthal modulation of X-rays scattered off a beryllium conus. The analyzing power is close to unity in agreement with expectations and Monte Carlo simulation results. Instrumental asymmetries of a few percent have been measured and corrected for with residual statistical uncertainties of O (10−3). The orientation of the polarization plane is determined to within less than 4 milliradians.

[1] J.They et al., SPIE vol 4138, 193 (2000)

100% | Mobil-Ansicht | English Version | Kontakt/Impressum/Datenschutz
DPG-Physik > DPG-Verhandlungen > 2001 > Erlangen