X-ray diagnostics and calibration of the COMPASS straw detectors — •Klaus Platzer1, Wolfgang Duennweber1, and Hermann Wellenstein2 — 1Sektion Physik der LMU, Am Coulombwall 1, 85748 Garching — 2Physics Department, Brandeis University, Waltham MA 02454
A scanning device consisting of a continuous beam, 50kV X-ray tube and a 20mm x 30mm CCD has been installed for the inspection of the straw tracking system of the COMPASS experiment at CERN. One double layer of about 800 straws, covering an area of 3.2m x 2.7m, is scanned in about 30 hours. The result is a grid of 5000 wire positions. The control of wire and wall spacing is possible with a local resolution better than 5 µm. The absolute wire positions are determined with an accuracy of about 50 µm, wich is sufficient as compared with the particle tracking resolution of about 200 µm of a single straw.