Bereiche | Tage | Auswahl | Suche | Aktualisierungen | Downloads | Hilfe
HK: Fachverband Physik der Hadronen und Kerne
HK 64: Instrumentierung
HK 64.7: Vortrag
Freitag, 21. März 2014, 15:30–15:45, HZ 9
Development of a Compton camera for online ion beam range verification via prompt γ detection * — •Saad Aldawood1,2, Christian Lang1, Hugh van der Kolff1,3, Ludwig Maier4, Rudolf Lutter1, Jonathan Bortfeldt1, Katia Parodi1, and Peter G. Thirolf1 — 1Ludwig-Maximilians-Universität München, Germany — 2King Saud University, Riyadh, Saudi Arabia — 3Delft University of Technology, The Netherlands — 4Technische Universität München, Germany
Precise and preferablly online ion beam range verification is a mandatory prerequisite to fully exploit the advantages of hadron-therapy in cancer treatment. Our aim is to develop an imaging system based on a Compton camera designed to detect prompt γ rays induced by nuclear reactions between ion beam and biological tissue. The Compton camera prototype consists of a stack of double-sided Si-strip detectors (DSSSD) acting as scatterers, while the absorber is formed by a LaBr3 scintillator crystal read out by a position-sensitive multi-anode photomultiplier. The LaBr3 detector was characterized with both absorptive and reflective side-face wrapping materials. Comparative studies of energy and time resolution, photopeak detection efficiency and spatial resolution will be presented together with first tests of the complete camera system.
* Supported by the DFG Cluster of Excellence MAP (Munich-Centre for Advanced Photonics).