Progress Report of an ultra-stable cavity for an Al+ optical clock — Sana Amairi Pyka1, •Kornelius Jakobsen1, Nils Scharnhorst1, Stephan Hannig1, Ian D. Leroux1, and Piet O. Schmidt1,2 — 1QUEST Institute, Physikalisch Technische Bundesanstalt, 38116 Braunschweig, Germany — 2Institut für Quantenoptik, Leibnitz-Universität Hannover, 30167 Hannover, Germany
We present the current status of the reference cavity for the clock laser of our aluminium ion frequency standard. A 1070 nm laser is stabilized to a 39.5 cm long cavity made of ultra-low expansion glass and is thereafter frequency-doubled twice to achieve the transition wavelength of 267 nm. The cavity has an estimated thermal noise limit of 7· 10−17 in 1 s. Despite its length, the system exhibits a very low sensitivity to vibrations of less than 10−11/m s−2 in all three directions. We present an evaluation of possible performance-limiting effects, such as residual amplitude noise and intensity stability of the laser light inside the cavity. These results are complemented by a comparison with other clock cavities at PTB. The evaluation shows that such long cavities are suitable for sub-Hertz spectroscopy at optical frequencies.