Thursday, March 17, 2016, 17:00–17:15, S1/05 23
THz Spectrometer Calibration at FELIX — •Toke Kövener, Steffen Wunderlich, Peter Peier, Eugen Hass, and Bernhard Schmidt — Deutsches Elektronen-Synchrotron, Notkestrasse 85, 22607 Hamburg, Germany
Coherent radiation spectroscopy is a suitable method for longitudinal electron bunch diagnostics at femtosecond bunch lengths. The absolute value of the longitudinal form factor, that is connected to the longitudinal profile, can be retrieved by measuring the intensity spectrum of a coherent transition radiation source at FLASH. The response function of the used spectrometer has to be well known in absolute values in order to perform accurate measurements. Until now, the response was predicted by calculations.
As the free-electron lasers at the FELIX facility in Nijmegen (NL) provide quasi-monochromatic beams that can be tuned in a wide spectral range at micrometer wavelengths, a calibration campaign for two THz spectrometers was performed at this facility with the goal to deduce their response function. Here we present the setup at FELIX that was used for the calibration scans, the achieved scan ranges and the collected data. Furthermore, the analysis of the measured data is discussed. The results are then compared to the previous calculations of the response functions.