Darmstadt 2016 – wissenschaftliches Programm
Bereiche | Tage | Auswahl | Suche | Aktualisierungen | Downloads | Hilfe
HK: Fachverband Physik der Hadronen und Kerne
HK 38: Instrumentation X
HK 38.5: Vortrag
Mittwoch, 16. März 2016, 15:15–15:30, S1/01 A2
High Resolution Time-of-Flight (TOF) Detector for Particle Identification — •Merlin Böhm, Albert Lehmann, Markus Pfaffinger, and Fred Uhlig for the PANDA collaboration — Physikalisches Institut, Universität Erlangen-Nürnberg
Several prototype tests were performed with the PANDA DIRC detectors at the CERN T9 beam line. A mixed hadron beam with pions, kaons and protons was used at momenta from 2 to 10 GeV/c. For these tests a good particle identification was mandatory.
We report about a high resolution TOF detector built especially for this purpose. It consists of two stations each consisting of a Cherenkov radiator read out by a Microchannel-Plate Photomultiplier (MCP-PMT) and a Scintillating Tile (SciTil) counter read out by silicon photomulipliers (SiPMs). With a flight path of 29 m a pion/kaon separation up to 5 GeV/c and a pion/proton separation up to 10 GeV/c was obtained. From the TOF resolutions of different counter combinations the time resolution (sigma) of the individual MCP-PMTs and SciTils was determined. The best counter reached a time resolution of 50 ps.
- Gefördert durch BMBF und GSI -