Mainz 2017 – wissenschaftliches Programm
Bereiche | Tage | Auswahl | Suche | Aktualisierungen | Downloads | Hilfe
MS: Fachverband Massenspektrometrie
MS 6: Accelerator Mass Spectrometry and Applications I
MS 6.7: Vortrag
Mittwoch, 8. März 2017, 16:15–16:30, RW 2
Quasi simultaneous measurements of 236U/238U ratios on the compact ETH Zurich Tandy AMS system — •Marcus Christl, Christof Vockenhuber, Sascha Maxeiner, Núria Casacuberta, Jürg Thut, and Hans-Arno Synal — Laboratory of Ion Beam Physics, ETH Zurich, Switzerland
Measurements of actinides with the compact 0.5 MV AMS system Tandy at ETH Zurich have become routine over the past years. With more than 200 samples analyzed per year (2012-2016), 236U currently represents the most important actinide nuclide measured at ETHZ. Very recently new, small Faraday Cups (FC) with an opening of 8x12 mm2 have been developed and installed at the HE side of the Tandy AMS system. With the new FCs it is now possible to perform quasi simultaneous measurements of 236U/238U ratios using the standard fast pulsing system. In this presentation the design of the FCs, the concept of the new AMS setup for 236U/238U and 236U/233U measurements, as well as the performance of the new FCs is presented. In addition to the technical part, new data from the ongoing oceanographic mapping program of 236U (and 129I) as well as the status of an intercalibration effort for 236U will be presented.