Bereiche | Tage | Auswahl | Suche | Aktualisierungen | Downloads | Hilfe
MS: Fachverband Massenspektrometrie
MS 9: Poster 1
MS 9.4: Poster
Donnerstag, 8. März 2018, 16:15–18:15, Redoutensaal
Concept of the automatic measurement using AMS setup at the Cologne FN Tandem Accelerator — •Susan Herb, Nima Saed-Samii, Claus Feuerstein, Markus Schiffer, Richard Altenkirch, Mario Cappellazzo, Stefan Heinze, Claus Müller-Gatermann, Gereon Hackenberg, and Alfred Dewald — Institut für Kernphysik, Universität zu Köln
The parameters of the hardware components of the AMS setup at the FN Tandem Accelerator are controlled by the in-house developed control software (phoenix software). In a first test phase AMS measurements at the FN Accelerator were performed manually using a standard amperemeter and the rare isotope data were processed by the commercial analog MPA3-system from FAST ComTec. In this contribution, we describe the concept and realization of an automated AMS-measurement. The timing for the sequential isotope injection is handled by an Arduino micro-controller which provides logical signals used to control the bouncer unit and as trigger for the acquisition. The detection signals are processed via a new in-house programmed digital DAQ-system based on CAEN digitizers. The phoenix software and the data acquisition are coupled over an Ethernet interface. Therefore the complete AMS data analysis is provided by the phoenix software.